FDA RECALL RECORD · Z-2625-2026

IntelliCuff, Portable automatic cuff pressure controller REF: 951001 Recall

IntelliCuff, Portable automatic cuff pressure controller REF: 951001

FDA recall Class II

Ongoing

FDA lists this Class II recall for IntelliCuff, Portable automatic cuff pressure controller REF: 951001. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2625-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 22, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
962 units
Distribution
See key details
FDA cause
Process design

What is being recalled?

IntelliCuff, Portable automatic cuff pressure controller REF: 951001

Affected model & UDI checker

Search the exact identifiers published for Z-2625-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
951001IntelliCuff, Portable automatic cuff pressure controller REF: 951001Check source noticeNot specified
Not specifiedIntelliCuff, Portable automatic cuff pressure controller REF: 951001Check source notice07630002800839
FDA code information

GTIN: 07630002800839/ Lot # range: 00001 -19732

View the original FDA recall notice

Why was this product recalled?

Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.

What actions did Hamilton Medical AG take?

On May 22, 2026, Hamilton Medical issued a Urgent Medical Device Recall to affected consignees. Hamilton Medical ask consignees to take the following actions: 1. Identify affected devices (consider devices in use, in stock, spare devices etc.) and make them available for exchange of cuff connectors. 2. Continue to use IntelliCuff devices according to their labelling. Mitigate potential risks by following the actions described in point B. Type of Action to mitigate the risk . 3. Please fill in and sign Customer Reply Form (page 5) and send it to your Hamilton Medical distribution partner or subsidiary. This must be conducted as quickly as possible, but no later than 30 calendar days after receiving the Field Safety Notice.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology