Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2638-2025

Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively). Recall

Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).

FDA recall Class II

Ongoing

FDA lists this Class II recall for Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2638-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 21, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
1,725
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).

Affected model & UDI checker

Search the exact identifiers published for Z-2638-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
VCRFG1Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).Check source noticeNot specified
Not specifiedVenclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).Check source notice00858254006022
FDA code information

UDI-DI: 00858254006022. Generator Software Version 3.35

View the original FDA recall notice

Why was this product recalled?

Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

What actions did Bard Peripheral Vascular, Inc. take?

On 8/21/2025, correction notices were mailed or emailed to Health Care Professionals (HCPs), Risk Manager, Head Nurse, Nurse Manager, Medical Device Safety Officer, Procurement Officer, Office Manager, and Distributors, were advised of the following: If the generator is operating with affected software and is connected to a catheter and the software check fails, a red "X" will be displayed on the generator screen without an accompanying error code Actions: 1) Firm Representative will contact your facility to arrange the restoration of your facility's generator to software version 3.17 upon receiving regulatory clearance of the 510(k) premarket notification. 2) Your facility will become eligible to receive future shipments of Venclose EVSRF Ablation Catheters only after the Venclose digiRF Generator has been verified to run software version 3.17. 3) Complete and return the customer response form via email to BDRC14@bd.com 4) Post the Customer Notification Letter on or near the generator until the software has been restored to software version 3.17 by a firm representative. 5) Share this notification with all product users within your facility network to ensure they are also aware of this product correction. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the recall. Please contact your firm representative if you require additional assistance with this process: North American Regional Complaint Center, 1-844-823-5433, productcomplaints@bd.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history