FDA RECALL RECORD · Z-2638-2026

Adapter for Head Positioning Recall

Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables

FDA recall Class II

Ongoing

FDA lists this Class II recall for Adapter for Head Positioning. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2638-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 21, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
13 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables

Affected model & UDI checker

Search the exact identifiers published for Z-2638-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1739994Adapter for Head PositioningCheck source noticeNot specified
Not specifiedAdapter for Head PositioningCheck source notice00887761972780
FDA code information

Lot Code: Product Code: 1739994 UDI-DI: 00887761972780 Serial Numbers: 109931797, 109931798, 109931799, 109943588, 109943590, 109948550, 109964963, 109964964, 109964969, 109964970, 109964971, 109980700, and 110006694

View the original FDA recall notice

Why was this product recalled?

The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.

What actions did Baxter Healthcare Corporation take?

An URGENT PRODUCT RECALL notification letter dated 4/20/26 was sent to customers. Actions to be Taken by Customers 1. Please notify all users of the product affected within your facility, especially all operating room staff. The affected Adapters may continue to be used only when mounted directly to the side rails of the operating table back plate. Do not use the Adapter on the Upper Back Section H V U side rails (component). Using the Upper Back Section H V U (separate component which is attached to the operating table s back plate) in conjunction with the Adapter accessory must be stopped/ceased, until a replacement takes place. 2. Once the newly designed Adapters are available Baxter will contact customers with instructions for return and replacement of the impacted Adapters. 3. If you received this communication directly from Baxter, acknowledge receipt by following the within the body of the corresponding email. Acknowledging receipt of this notification will prevent you from receiving repeated notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please contact your supplier to arrange for return of the affected product. Please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to the director of nursing, facility risk manager, patient safety, biomedical engineering department, director of surgical services, operating room staff, and any other departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.