Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2645-2026

Brand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN Recall

Brand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN Model/Catalog Number: ENF060182LD / 101033086 Software Version: N/A Product Description: Enteral Extension Set, Length 60 Orange-Striped Minibore Tubing, (Approx. Priming Vol. 2.00 mL), ENFit¿ Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2645-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 5, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
82900
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Brand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN Model/Catalog Number: ENF060182LD / 101033086 Software Version: N/A Product Description: Enteral Extension Set, Length 60 Orange-Striped Minibore Tubing, (Approx. Priming Vol. 2.00 mL), ENFit Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-2645-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
ENF060182LDBrand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct ENCheck source noticeNot specified
Not specifiedBrand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct ENCheck source notice20020451101358
Not specifiedBrand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct ENCheck source notice00020451101354
FDA code information

Lot Code: Item Number ENF060182LD / 101033086; Lot Number 0000747253; Date of Manufacturing 2026-01-16; Expiration Date 2029-01-16; Carton GTIN 20020451101358; Unit GTIN 00020451101354

View the original FDA recall notice

Why was this product recalled?

May contain incorrect connectors and will not attach or misconnect to ENFit connections and is thus incompatible. Issue could lead to slight delay in delivery of enteral nutritional fluids or oral medication.

What actions did Medela Inc take?

On June 5, 2026, the firm notified customers via letters titled "URGENT Medical Device Recall." Customers were instructed to quarantine affected product lots, dispose of products, and complete and return a customer response form. Upon receipt of the completed response form, Medela will contact the customer to replace product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology