FDA RECALL RECORD · Z-2648-2026

ARTIS icono ceiling. Recall

ARTIS icono ceiling. Model Number: 11328100.

FDA recall Class II

Ongoing

FDA lists this Class II recall for ARTIS icono ceiling.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2648-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 8, 2026
FDA report date
Jul 8, 2026
Quantity in commerce
57 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

ARTIS icono ceiling. Model Number: 11328100.

Affected model & UDI checker

Search the exact identifiers published for Z-2648-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11328100ARTIS icono ceiling.Check source noticeNot specified
Not specifiedARTIS icono ceiling.Check source notice4056869295923
FDA code information

Model Number: 11328100. UDI-DI: 4056869295923.

View the original FDA recall notice

Why was this product recalled?

Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Medical Solutions USA, Inc. take?

Siemens Healthineers notified consignees on about 06/08/2026 via "CUSTOMER SAFETY ADVISORY NOTICE" letter. Consignees were instructed to conduct regular inspections of the gap between the monitor carrier and the ceiling-mounted arm and be attentive to any unusual noises, shaking, wobbling, or other unexpected behavior during system operation, as these may indicate loose screws. If any such signs are observed, limit movements of the DCS and contact Siemens service organization immediately for further instructions. Always ensure that the DCS is positioned for safe use. All affected systems will be corrected by Siemens Healthineers service engineers. Siemens engineers will correct the screwed connection of the monitor ceiling support according to the adapted specifications and instructions. The correction will be released via an Update Instruction AX022/26/S. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.