Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2662-2025

MAGNETOM Skyra Fit BioMatrix. Recall

MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.

FDA recall Class I

Ongoing

FDA lists this Class I recall for MAGNETOM Skyra Fit BioMatrix.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2662-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Aug 28, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
39 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.

Affected model & UDI checker

Search the exact identifiers published for Z-2662-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11516217MAGNETOM Skyra Fit BioMatrix.Check source noticeNot specified
Not specifiedMAGNETOM Skyra Fit BioMatrix.Check source notice04056869299273
FDA code information

Model Number: 11516217. UDI Numbers: (01)04056869299273(21)212087, (01)04056869299273(21)212098, (01)04056869299273(21)212088, (01)04056869299273(21)212068, (01)04056869299273(21)212081, (01)04056869299273(21)212018, (01)04056869299273(21)212041, (01)04056869299273(21)212038, (01)04056869299273(21)212047, (01)04056869299273(21)212019, (01)04056869299273(21)212037, (01)04056869299273(21)212032, (01)04056869299273(21)212049, (01)04056869299273(21)212017, (01)04056869299273(21)212064, (01)04056869299273(21)212096, (01)04056869299273(21)212065, (01)04056869299273(21)212016, (01)04056869299273(21)212023, (01)04056869299273(21)212056, (01)04056869299273(21)212054, (01)04056869299273(21)212063, (01)04056869299273(21)212051, (01)04056869299273(21)212055, (01)04056869299273(21)212021, (01)04056869299273(21)212060, (01)04056869299273(21)212011, (01)04056869299273(21)212066, (01)04056869299273(21)212074, (01)04056869299273(21)212090, (01)04056869299273(21)212034, (01)04056869299273(21)212029, (01)04056869299273(21)212097, (01)04056869299273(21)212099, (01)04056869299273(21)212039, (01)04056869299273(21)212059, (01)04056869299273(21)212030, (01)04056869299273(21)212035, (01)04056869299273(21)212036. Serial Numbers: 212087, 212098, 212088, 212068, 212081, 212018, 212041, 212038, 212047, 212019, 212037, 212032, 212049, 212017, 212064, 212096, 212065, 212016, 212023, 212056, 212054, 212063, 212051, 212055, 212021, 212060, 212011, 212066, 212074, 212090, 212034, 212029, 212097, 212099, 212039, 212059, 212030, 212035, 212036.

View the original FDA recall notice

Why was this product recalled?

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Medical Solutions USA, Inc. take?

Siemens Healthcare notified consignees on about 08/28/2025 via letter. Consignees were informed that Siemens Customer Service Organization will conduct an inspection of affected units to address the issue. If customers identify issues, they are cautioned to not reboot the system, to call their local Siemens Healthineers service organization, and grant access to the system. Siemens will inspect devices and perform a correction, if needed. Consignees were also instructed to ensure all users or customers are notified of the issue and to maintain the Customer Safety Advisory Notice. Consignees were also requested to confirm receipt and acknowledgement of the notification. Siemens expanded the recall and notified consignees on about 10/01/2025 via letter. These notifications included the same information as initially sent to consignees. A follow up communication was sent to consignees on about 10/15/2025 via letter. This informed users of further details and additional necessary safety actions. If a specific "Magnent Supervision" error message appears on the host computer, no further scans can be performed, and you are requested to: not reboot the system, remove the patient from the system, lock/secure all access to the scanning room with the system remaining energized at field, post a DO NOT USE notice on or near the device, and call the Siemens Healthineers service organization. In addition, Siemens reminded users that the Magnet Stop button is only intended for emergencies. DO NOT initiate a manual quench of the magnet unless there is an emergency. Ensure all designated emergency exit routes remain unobstructed and are marked. Consignees were also instructed to ensure all users or customers are notified of the issue, to maintain the Customer Safety Advisory Notice is retained in product related records, to confirm receipt and acknowledgement of the notification, and if the affected system is no longer in the consignee's possession, notify the new owner. ***Updat

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology