Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System) Recall
Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System)
FDA recall Class II
Ongoing
FDA lists this Class II recall for Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2673-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 9, 2025
- FDA report date
- Oct 8, 2025
- Quantity in commerce
- 474 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System)
Affected model & UDI checker
Search the exact identifiers published for Z-2673-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System) | Check source notice | 4260648620046 |
FDA code information
Model Name: Ashvins variant HEYEX 2 / HEYEX PACS. UDI-DI: 4260648620046. Software Version: 2.6.0 (Build 2088) to 2.6.8 (Build 2220).
Why was this product recalled?
Potential that the measured value may be smaller than the actual area.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did MedicalCommunications GmbH take?
Ashvins notified consignees on about 09/09/2025 via emailed letter. Consignees were instructed to cease using the freeform measurement function in the MMV until the software update has been performed, which is now available. Consignees were instructed to perform the update as soon as possible, complete the digital Field Safety Notice Acknowledgement and Receipt Form, and maintain a copy of the letter and the user manual with the device until the software update is complete.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2673-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026