MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 Recall
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
FDA recall Class II
Ongoing
FDA lists this Class II recall for MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2698-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 27, 2026
- FDA report date
- Jul 15, 2026
- Quantity in commerce
- 778
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Affected model & UDI checker
Search the exact identifiers published for Z-2698-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MMT-1885 | MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 | Check source notice | Not specified |
| MMT-1886 | MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 | Check source notice | Not specified |
| MMT-1895 | MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 | Check source notice | Not specified |
| MMT-1896 | MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 | Check source notice | Not specified |
FDA code information
UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
Why was this product recalled?
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
What actions did Medtronic MiniMed, Inc. take?
On 5/27/2026, field safety notices were emailed to customers who were asked to do the following: 1) Update insulin pump software to version 6.63. The update may take up to 3 hours to complete. 2) Reply to this email to confirm that you have read and understood this notification and, where applicable, completed the required action. If you have questions, contact the firm at 01923 205167
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2698-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.