Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation. Recall
Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2735-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Jun 25, 2026
- FDA report date
- Aug 5, 2026
- Quantity in commerce
- 168,069 devices globally
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.
Affected model & UDI checker
Search the exact identifiers published for Z-2735-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation. | Check source notice | 00619498003259 |
| Not specified | Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation. | Check source notice | 00619498270033 |
FDA code information
Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070 Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070
Why was this product recalled?
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
What actions did ResMed Ltd. take?
ResMed Urgent Medical Device Correction Astral 100/150 Ventilators Issue Identified: On June 25, 2026, ResMed notified customers via Salesforce of an urgent medical device correction involving a subset of Astral 100 and Astral 150 ventilators. An internal supercapacitor may leak electrolyte over time, potentially damaging the printed circuit board assembly (PCBA) and causing the ventilator to enter a fail-safe state. Interim Guidance: Due to limited PCBA replacement availability, immediate correction of all affected units is not feasible. Physicians and patients are advised to follow Astral 100/150 User and Clinical Guides, including recommendations on monitoring, emergency preparedness, caregiver training, and backup ventilation. Required Actions for Healthcare Providers and Distributors: Acknowledge: Complete and return the acknowledgment form by July 31, 2026. Notify: Distribute the correction notice and accompanying letters to all relevant healthcare professionals, patients, and caregivers immediately. Reinforce Safety Measures: Ensure ventilator-dependent patients are monitored, caregivers are trained in alarm response and emergency procedures, and backup ventilation equipment is functional and readily available. Device Management: Do not remove an affected ventilator from use unless an alternative is immediately available. Replace inoperable units promptly; patients should continue therapy unless directed otherwise by their clinician. Risk Assessment: Evaluate affected patients by clinical risk category (Tier 1, 2, or 3) per Appendix A. Arrange Inspection: Schedule ventilator inspections per the Phase 1 Inspection and Correction Strategy in Appendix A, returning units to an authorized service center as needed. Contact: Reach your local ResMed representative or visit www.resmed.com/contact for questions.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2735-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology