Power Over Fiber Base Module Recall
Power Over Fiber Base Module
FDA recall Class II
Ongoing
FDA lists this Class II recall for Power Over Fiber Base Module. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2745-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 11, 2026
- FDA report date
- Aug 5, 2026
- Quantity in commerce
- 101
- Distribution
- See key details
- FDA cause
- Radiation Control for Health and Safety Act
What is being recalled?
Power Over Fiber Base Module
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Power Over Fiber Base Module Identification P/N 633232 P/N 622365 P/N 633383 P/N 633910 P/N 635731
Why was this product recalled?
Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement
What actions did Terex Utilities take?
A letter was received stating the product, problem and actions to be taken. The following actions being taken include: A. Develop and carry out a process to remove incorrect labeling and install all labels required under IEC 60825-1 2014. B. Generate, detail & submit full CAP (Corrective Action Plan) for FDA approval C. Generate initial laser product report or supplemental as appropriate. D. Equipment update to be performed by Terex personnel at customer location E. Procedures: * Fully certify device as Class 1 under the guidelines of FDA Laser Notice #56 by applying the IEC 60825-1 2014 laser certification standard * Update and distribute compliant UI (User Information) * Update Sales information as appropriate * Generate and apply administrative and safety labeling * Formalized quality control procedures and record keeping. If you have any questions, contact Director of Engineering at +16058824000 or email: dan.brenden@terex.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2745-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology