Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2750-2026

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter Recall

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter

FDA recall Class II

Ongoing

FDA lists this Class II recall for FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2750-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 11, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
18 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter

Affected model & UDI checker

Search the exact identifiers published for Z-2750-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedFARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation CatheterCheck source notice00810087180126
FDA code information

GTIN 00810087180126, Serial Numbers: 102174748, 103355157, 103432386, 103480452, 103547915, 103836571, 1-215-0040, 1-226-0123, 1-227-0129, 1-228-0139, 1-22C-0216, 1-234-0276, 1-237-0307, 104015648, 104993721, 1-233-0266, 1-236-0305, 1-237-0307.

View the original FDA recall notice

Why was this product recalled?

Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.

What actions did BOSTON SCIENTIFIC CORPORATION take?

Boston Scientific issued an Urgent Field Safety Notice - Urgent Medical Device Correction notice to its consignees out side the US on 6/11/2026 via hand delivered letter. The notice explained the issue and requested the following actions be taken: Instructions: 1- Post a copy of the appropriate IFU provided near each FARASTAR PFA generator, replacing any previous versions currently in use. 2- Review the IFU and ensure all relevant personnel are aware of its contents. 3- Please complete the attached Acknowledgement Form even if you do not have any affected product and return it to your Boston Scientific office. 4- Share this communication with any healthcare professionals in your facility that use the product and with any other organization to which this product may have been transferred. The national Competent Authority was informed of this communication. Any adverse events or quality concerns associated with use of these devices should be reported to Boston Scientific and Competent Authorities if appropriate.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology