Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2753-2026

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter Recall

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter

FDA recall Class II

Ongoing

FDA lists this Class II recall for FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2753-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 11, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
6 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter

Affected model & UDI checker

Search the exact identifiers published for Z-2753-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedFARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation CatheterCheck source notice00810087180126
FDA code information

GTIN/UDI 00810087180126, Serial Numbers: 1-217-0053, 1-21C-0078, 1-223-0099, 1-225-0106, 1-225-0107, 1-229-0172.

View the original FDA recall notice

Why was this product recalled?

Boston Scientific is replacing specific FARASTAR PFA generators which have a configuration that is not compatible for use with the FARAPOINT PFA catheter.

What actions did BOSTON SCIENTIFIC CORPORATION take?

Boston Scientific issued an Urgent Field Safety Notice - Urgent Medical Device Recall notice to its consignees out side the US on 6/11/2026 via hand delivered letter. The notice explained the issue and requested the following actions be taken: Instructions: 2- Allow Boston Scientific representative(s) to replace affected FARASTAR PFA generators. 3- Please complete the attached Verification Form even if you do not have any product to return and send it to your local Boston Scientific office 4- If you have products to return, please package them in an appropriate shipping box. After receipt of the Verification Form, Boston Scientific will contact you to arrange return. 6- Please pass this notice to any health professional from your organization that needs to be aware, and to any organization where the potentially affected devices have been transferred (if appropriate). Please provide Boston Scientific with details of any affected devices that have been transferred to other organizations (if appropriate). The National Competent Authority has been informed of this communication.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology