Olympus 4K UHD LCD Monitor. Recall
Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Olympus 4K UHD LCD Monitor.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2761-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 7, 2026
- FDA report date
- Jul 29, 2026
- Quantity in commerce
- 10 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.
Affected model & UDI checker
Search the exact identifiers published for Z-2761-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| OEV321UH | Olympus 4K UHD LCD Monitor. | Check source notice | Not specified |
| Not specified | Olympus 4K UHD LCD Monitor. | Check source notice | 04953170418723 |
FDA code information
Model Number: OEV321UH. UDI-DI: 04953170418723. Serial Numbers: 7503474, 7503477, 7503479, 7503263, 7503264, 7503262, 7503261, 7503260, 7503259, 7503335.
Why was this product recalled?
Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Olympus Corporation of the Americas take?
On July 7, 2026, the firm issued "Urgent: Medical Device Correction" letters to consignees. Customers were informed of the product issue. While the identified 4K UHD LCD monitor may be used with VISERA S, VISERA ELITE III, EVIS X1, and EU-ME3 systems, the firm notes that the affected HDMI or DP connectors are more commonly used with the VISERA S video system center. Its use with the three other systems does not typically involve the affected HDMI or DP connectors. Customers are requested to examine their equipment, looking for the identified monitor model number and serial numbers listed in the customer letter. They should then acknowledge receipt of the letter through the firm's recall web portal. The firm will coordinate repair of the monitors. If customer needs to use the monitor prior to its repair, the firm recommends that a spare unit be available in case issues are encountered with the primary unit during use.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2761-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology