FDA RECALL RECORD · Z-2761-2026

Olympus 4K UHD LCD Monitor. Recall

Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Olympus 4K UHD LCD Monitor.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2761-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 7, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
10 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.

Affected model & UDI checker

Search the exact identifiers published for Z-2761-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
OEV321UHOlympus 4K UHD LCD Monitor.Check source noticeNot specified
Not specifiedOlympus 4K UHD LCD Monitor.Check source notice04953170418723
FDA code information

Model Number: OEV321UH. UDI-DI: 04953170418723. Serial Numbers: 7503474, 7503477, 7503479, 7503263, 7503264, 7503262, 7503261, 7503260, 7503259, 7503335.

View the original FDA recall notice

Why was this product recalled?

Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Olympus Corporation of the Americas take?

On July 7, 2026, the firm issued "Urgent: Medical Device Correction" letters to consignees. Customers were informed of the product issue. While the identified 4K UHD LCD monitor may be used with VISERA S, VISERA ELITE III, EVIS X1, and EU-ME3 systems, the firm notes that the affected HDMI or DP connectors are more commonly used with the VISERA S video system center. Its use with the three other systems does not typically involve the affected HDMI or DP connectors. Customers are requested to examine their equipment, looking for the identified monitor model number and serial numbers listed in the customer letter. They should then acknowledge receipt of the letter through the firm's recall web portal. The firm will coordinate repair of the monitors. If customer needs to use the monitor prior to its repair, the firm recommends that a spare unit be available in case issues are encountered with the primary unit during use.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.