Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2766-2026

Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets Recall

Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets

FDA recall Class I

Ongoing

FDA lists this Class I recall for Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2766-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jun 30, 2026
FDA report date
Aug 19, 2026
Quantity in commerce
24528 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets

Affected model & UDI checker

Search the exact identifiers published for Z-2766-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedBaxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration SetsCheck source notice00085412014319
FDA code information

UDI/DI 00085412014319, Lot Numbers: DR26C27062, exp. 30-MAR-2028; DR26D14031, exp. 15-Apr-2028; DR26E13055, exp. 14-May-2028

View the original FDA recall notice

Why was this product recalled?

Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.

What actions did Baxter Healthcare Corporation take?

Baxter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/30/2026 via USPS, first class mail. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Immediately locate, isolate, and cease all use of the affected lot numbers of the product. The product code and lot number can be found on the individual product package and shipping carton. 2. Contact Baxter Healthcare Center for Service to arrange for return and replacement product at 888-229-0001 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you received this communication directly from Baxter, please acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please contact them to arrange for return of the affected product. Please note that responding to Baxter is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to the director of nursing, director of pharmacy, director of purchasing, facility risk manager, and any other departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of the affected product that you distributed to customers and check the associated box on

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology