Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Recall
Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2769-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 27, 2026
- FDA report date
- Jul 29, 2026
- Quantity in commerce
- 20
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Item Number: NITESL UDI-DI code: D745SILNT0 Lot/Serial Numbers: 1020214577 1020784354 1030007997 1030439672 1030523763 1030722759 1106770735 1107310424 1107642834 1108530251 1110784932 1117400787 1118814629 1121913952 1122251905 1122958036 1124408531 1125495087 1127571733 1131149684
Why was this product recalled?
Due to device not having FDA Clearance
What actions did Prismatik Dentalcraft, Inc. take?
On 04/27/2026, the firm sent via FedEx an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL" Letter to customers informing them that Prismatik discovered that a subset of Silent Nite (Product SKU Number: SILNT, Product ID: NITESL) devices were manufactured using Erkoflex-95, a material not cleared by the FDA for manufacturing the Silent Nite. Customers are instructed to: 1. What should Customers do with the product? For the product listed above, please discontinue use and if you have the product with you, please return it to receive a credit, using the return form attached with the specific contact information to: RA.Mailbox@glidewelldental.com. 2. Notify all relevant personnel within your organization and all patients to whom the affected products may have been prescribed for this recall and the information contained in this letter. 3. When completing the form, indicate the quantities that need to be returned. Please sign, date, and return the form to RA.Mailbox@glidewelldental.com. Keep your quarantine product segregated; our Customer Service Representative will be contacting you with further instructions. 4. Please make a copy of the signed form and place it with the product being returned. This will identify the product as Recall Product for proper handling and disposition. 5. If you do not have product to be returned, place an X next to the statement, No Product to Return , sign, date and return the form to RA.Mailbox@glidewelldental.com. For questions - contact Regulatory Affairs at (949) 836-5479 or email : RA.Mailbox@glidewelldental.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2769-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology