Olympus High Flow Insufflation Unit - UHI-4. Recall

Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4

FDA recall Class II

Ongoing

FDA lists this Class II recall for Olympus High Flow Insufflation Unit - UHI-4.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2776-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 16, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
3,087 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4

Affected model & UDI checker

Search the exact identifiers published for Z-2776-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
UHI-4.Olympus High Flow Insufflation Unit - UHI-4.Check source notice04953170324147
UHI-4Olympus High Flow Insufflation Unit - UHI-4.Check source noticeNot specified
Not specifiedOlympus High Flow Insufflation Unit - UHI-4.Check source notice04953170467219
FDA code information

Model Number: UHI-4. UDI: 04953170324147, 04953170467219. All Serial Numbers.

View the original FDA recall notice

Why was this product recalled?

Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on about 06/16/2026 via letter. Consignees were informed that they many continue to use their UHI-4 until the offset adjustment is performed. However, as noted in the Instructions for Use (IFU), prepare a spare UHI-4 unit as a backup to ensure uninterrupted operation can be completed without complication in the case of a malfunction. Actions to take included to carefully read the notification letter and ensure all personnel are completely knowledgeable on the content of the notification, an Olympus field service representative will contact you to schedule a mutually convenient time to perform the offset adjustment, acknowledge receipt of the notification through the Olympus recall web portal, and forward the notification to other users who may have the affected products if you have further distributed it.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology