Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application. Recall
Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2777-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 29, 2026
- FDA report date
- Jul 29, 2026
- Quantity in commerce
- 685
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.
Affected model & UDI checker
Search the exact identifiers published for Z-2777-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application. | Check source notice | 199874903484 |
FDA code information
Software Version - 2.0.0 GS1 GTIN code: 199874903484 UDI code: GUDID registration is in progress
Why was this product recalled?
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
What actions did Emano Metrics, Inc. take?
On 06/10/2026, a "Medical Device Correction" Letter was emailed to customers informing them that it would found that Emano Flow software versions v2.2.0-v2.3.4 processed audio at a sample rate that did not match the sample rate expected by the flow prediction algorithms (model version v2.0.0), affecting results generated between 03/22/2026 to 05/29/2026. Customer were also informed that, the correction (updated software) was deployed on May 29, 2026. An updated customer notice was emailed on 06/23/2026. Instructions provided to Customer included: 1. Review the corrected results for patients with Emano Flow data from the affected period, prioritizing cases where the data informed a clinical decision. 2. If customers have previously downloaded patient reports generated between 03/22/2026 and 05/29/2026, replace them with the corrected versions available in the Emano Flow clinic portal. A spreadsheet with secure links to all corrected reports is attached. 3. Please reply to this email to confirm receipt of this notice.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2777-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data