Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A. Recall
Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2788-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 11, 2026
- FDA report date
- Jul 29, 2026
- Quantity in commerce
- 155 kits
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.
Affected model & UDI checker
Search the exact identifiers published for Z-2788-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A. | Check source notice | 10198459346873 |
| Not specified | Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A. | Check source notice | 40198459346874 |
| Not specified | Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A. | Check source notice | 10198459213342 |
| Not specified | Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A. | Check source notice | 40198459213343 |
FDA code information
Medline Kit SKU DYNJ910103B, UDI/DI 10198459346873 (each) 40198459346874 (case), Lot Number 25FBR237; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25ABM761; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25BBD107; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBD534; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBL772.
Why was this product recalled?
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
What actions did Medline Industries, LP take?
Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees via email and USPS first class mail. The notice explained the issue with the device, potential harms, and requested the following: REQUIRED ACTION: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete the response form. Please list the quantity of affected product in inventory on the form. Even if there is no affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com 3. Upon receipt of the response form, the consignee account will receive over-labels to place on affected inventory, with instructions for the staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors and those who transferred the affected product were directed to notify those to which the product was distributed or transferred. For contact the Recall Department at 866-359-1704 or recalls@medline.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2788-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology