Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2792-2026

Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds Recall

Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds

FDA recall Class I

Ongoing

FDA lists this Class I recall for Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2792-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
May 13, 2026
FDA report date
Aug 5, 2026
Quantity in commerce
657
Distribution
See key details
FDA cause
Material/Component Contamination

What is being recalled?

Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds

Affected model & UDI checker

Search the exact identifiers published for Z-2792-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
NCAG62IPack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle ManifoldsCheck source noticeNot specified
Not specifiedPack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle ManifoldsCheck source notice20809160490416
FDA code information

UDI-DI: 20809160490416. Lot/Expiration: 342513(1/16/2028), 336758(9/20/2027), 338977(1/16/2028)

View the original FDA recall notice

Why was this product recalled?

Cardiovascular procedure kits contain manifolds, the subject of another recall, that may have particulate within the fluid path, which has the potential to be introduced into blood circulation and become lodged within blood vessels.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medical Action Industries, Inc. 306 take?

On 5/13/2026, Owens & Minor emailed recall notices to customers who were asked to do the following: Identify, segregate, and quarantine all impacted product. Other components within the convenience kits can still be used once kit is properly labeled with the warning label below. Add warning labels to affected Medical Action kits to ensure all users are aware of the actions to remove and discard Manifold component as well as actions identified below and in the enclosed Medline recall notice. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Labels state in part: Prior to use, remove & discard the Medline Namic Angiographic (The Manifold). Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. Complete and return the response form via email to GM-OMRA_RECALLS@owens-minor.com Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology