Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2794-2026

Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1. Recall

Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2794-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 21, 2026
FDA report date
Jul 29, 2026
Quantity in commerce
45
Distribution
See key details
FDA cause
Packaging change control

What is being recalled?

Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.

Affected model & UDI checker

Search the exact identifiers published for Z-2794-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11264BBU1Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.Check source noticeNot specified
26153EAU/1Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.Check source noticeNot specified
Not specifiedFlexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.Check source notice04048551226087
FDA code information

Kit UDI-DI: 04048551226087. Kit Serial Numbers: 2274269, 2273282, 2276217, 2276216, 2264656, 2275787, 2279756, 2280154, 2280379, 2275789, 2283453, 2277877, 2280204, 2273324, 2275986, 2283483, 2229591, 2273599, 2215800, 2274197, 2280156, 2275483, 2083136, 2275790, 2275482, 2274271, 2248338, 2283444, 2188673, 2276221, 2276217, 2275788, 2274199, 2283445, 2274198, 2274041, 2283716, 2274049, 2236228, 2279757, 2274267, 2275785, 2277894, 2230599, 2267785. 26153EAU/1 seals with expiration date: January 23, 2023.

View the original FDA recall notice

Why was this product recalled?

Flexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal for gynecology use, which could impact the assurance of sterility.

What actions did Karl Storz Endoscopy take?

On 5/21/2026, correction notices were emailed to customers who were asked to do the following: 1) Check the expiration date for any inventory of 26153EAU-01 Sealing Seal for Gynecology and discard any expired components(seals) by removing the seal from the packaging and discarding the seal and packaging. 2) Pass on this Urgent Medical Device Correction to all users of affected products and all other persons who need to be aware within your organization. 3) If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice and response form. 4) Completed and return the Customer Reply Form via email to KSEAProductLabeling@karlstorz.com For all other questions regarding this communication, please send an email to KSEAProductLabeling@karlstorz.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.