Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05 Recall
Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05
FDA recall Class I
Ongoing
FDA lists this Class I recall for Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2796-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Apr 30, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 5168 kits
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05
Affected model & UDI checker
Search the exact identifiers published for Z-2796-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05 | Check source notice | 10809160377314 |
FDA code information
UDI-DI 10809160377314 Lots 1573593 1583827 1609541 1612134 1612516 1619132 1623384 1627013 1635459 1672306 1675159
Why was this product recalled?
Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did AVID Medical, Inc. take?
On April 30, 2026, firm notified distributors via email. Distributors were tasked with notifying their end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. They were provided with a labeling template to print warning labels which should be affixed on all affected kits and cases in inventory. The firm is also able to provide pre-printed labels if needed. Warning labels inform users to remove and discard the manifold prior to use.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2796-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.