Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2797-2026

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15 Recall

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15

FDA recall Class I

Ongoing

FDA lists this Class I recall for Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2797-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Apr 30, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
5168 kits
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15

Affected model & UDI checker

Search the exact identifiers published for Z-2797-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedHalyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15Check source notice10809160388495
FDA code information

UDI-DI 10809160388495 Lots 1561685 1575926 1583048 1584017 1589652 1595715 1599634 1605749 1611810 1621476 1637222 1637224 1637225 1643118 1645143 1645637 1663929 1663930 1663931 1663932

View the original FDA recall notice

Why was this product recalled?

Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AVID Medical, Inc. take?

On April 30, 2026, firm notified distributors via email. Distributors were tasked with notifying their end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. They were provided with a labeling template to print warning labels which should be affixed on all affected kits and cases in inventory. The firm is also able to provide pre-printed labels if needed. Warning labels inform users to remove and discard the manifold prior to use.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology