AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Laminectomy Pack, Pack Number AMS13574A; 2) ACDF/Fusion Pack, Pack Number AMS14710B; 3) Spine Pack, Pack Number AMS14991A; 4) Spine Pack, Pack Number AMS14991B; 5) Spinal Pack, Pack Number AMS16278; 6) Spine Pack, Pack Number AMS16499; 7) Spine Pack, Pack Number AMS16709; 8) Lumbar Spine Pack, Pack Number AMS17336; 9) Laminectomy Pack, Pack Number AMS17945; 10) Anterior Cervical Pack, Pack Number AMS17946; 11) Spine Pack, Pack Number AMS3150F; 12) Spine Pack, Pack Number AMS3150G; 13) Spine Pack, Pack Number AMS5451D; 14) ENT Pack, Pack Number AMS9771. Recall
AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Laminectomy Pack, Pack Number AMS13574A; 2) ACDF/Fusion Pack, Pack Number AMS14710B; 3) Spine Pack, Pack Number AMS14991A; 4) Spine Pack, Pack Number AMS14991B; 5) Spinal Pack, Pack Number AMS16278; 6) Spine Pack, Pack Number AMS16499; 7) Spine Pack, Pack Number AMS16709; 8) Lumbar Spine Pack, Pack Number AMS17336; 9) Laminectomy Pack, Pack Number AMS17945; 10) Anterior Cervical Pack, Pack Number AMS17946; 11) Spine Pack, Pack Number AMS3150F; 12) Spine Pack, Pack Number AMS3150G; 13) Spine Pack, Pack Number AMS5451D; 14) ENT Pack, Pack Number AMS9771.
FDA recall Class II
Ongoing
FDA lists this Class II recall for AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Laminectomy Pack, Pack Number AMS13574A; 2) ACDF/Fusion Pack, Pack Number AMS14710B; 3) Spine Pack, Pack Number AMS14991A; 4) Spine Pack, Pack Number AMS14991B; 5) Spinal Pack, Pack Number AMS16278; 6) Spine Pack, Pack Number AMS16499; 7) Spine Pack, Pack Number AMS16709; 8) Lumbar Spine Pack, Pack Number AMS17336; 9) Laminectomy Pack, Pack Number AMS17945; 10) Anterior Cervical Pack, Pack Number AMS17946; 11) Spine Pack, Pack Number AMS3150F; 12) Spine Pack, Pack Number AMS3150G; 13) Spine Pack, Pack Number AMS5451D; 14) ENT Pack, Pack Number AMS9771.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2802-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 8, 2026
- FDA report date
- Aug 5, 2026
- Quantity in commerce
- 7306 units
- Distribution
- See key details
- FDA cause
- Material/Component Contamination
What is being recalled?
AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Laminectomy Pack, Pack Number AMS13574A; 2) ACDF/Fusion Pack, Pack Number AMS14710B; 3) Spine Pack, Pack Number AMS14991A; 4) Spine Pack, Pack Number AMS14991B; 5) Spinal Pack, Pack Number AMS16278; 6) Spine Pack, Pack Number AMS16499; 7) Spine Pack, Pack Number AMS16709; 8) Lumbar Spine Pack, Pack Number AMS17336; 9) Laminectomy Pack, Pack Number AMS17945; 10) Anterior Cervical Pack, Pack Number AMS17946; 11) Spine Pack, Pack Number AMS3150F; 12) Spine Pack, Pack Number AMS3150G; 13) Spine Pack, Pack Number AMS5451D; 14) ENT Pack, Pack Number AMS9771.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
1) Pack Number AMS13574A, UDI B098AMS13574A0, Lot Numbers: 235911, 236897. 2) Pack Number AMS14710B, UDI B098AMS14710B0, Lot Numbers: 224717, 230666, 213914, 236351. 3) Pack Number AMS14991A, UDI B098AMS14991A0, Lot Numbers: 215450, 230108. 4) Pack Number AMS14991B, UDI B098AMS14991B0, Lot Numbers: 233346. 5) Pack Number AMS16278, UDI B098AMS162780, Lot Numbers: 219872, 222999, 226497, 229309, 236465, 237544, 238613. 6) Pack Number AMS16499, UDI B098AMS164990, Lot Numbers: 215711, 218234, 224253, 226168, 226627, 227667, 228446, 230422, 233097, 233399, 233960, 234577, 234907, 236220, 236852, 237358, 238563, 239114, 240373, 241010, 241717. 7) Pack Number AMS16709, UDI B098AMS167090, Lot Numbers: 223361, 225039, 236666, 240344, 241412. 8) Pack Number AMS17336, UDI B098AMS173360, Lot Numbers: 233541, 240967. 9) Pack Number AMS17945, UDI B098AMS179450, Lot Numbers: 235478, 240779, 241565. 10) Pack Number AMS17946, UDI B098AMS179460, Lot Numbers: 235954, 239092, 241514. 11) Pack Number AMS3150F, UDI B098AMS3150F0, Lot Numbers: 211888, 216076, 216357, 218443, 219827, 221799, 223766, 227324, 228151, 229404, 229717, 230623, 231694, 233211, 234586, 234763, 236581, 241332. 12) Pack Number AMS3150G, UDI B098AMS3150G0, Lot Numbers: 242374. 13) Pack Number AMS5451D, UDI B098AMS5451D0, Lot Numbers: 222621, 224568, 238089, 238575, 242113. 14) Pack Number AMS9771, UDI B098AMS97710, Lot Numbers: 176784, 179837, 181956, 184890, 191784, 194017, 196708, 197543, 199137, 201956, 210134, 215434, 217352, 218696, 225332, 226657, 228224, 228678, 230472, 233592, 235614, 236684, 241927.
Why was this product recalled?
The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Windstone Medical Packaging, Inc. take?
Aligned Medical Solutions (AMS) issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 6/8/2026 via email. The notice explained the issue, potential risk, and requested the following actions: identify and quarantine affected product, affix the sticker which reads: "RECALL NOTICE Medicom announced the recall of Ritmed Neurological Sponges contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall." " At the time the kit is opened for use any Medicom-Ritmed Neurological Sponges should be identified and set aside. " The recalled Medicom-Ritmed Neurological Sponges should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Complete the recall reply form and return it via email to: fieldcorrectiveaction@alignedmedical.com even if you have no affected product on hand. Those further distributing the product were directed to notify their distribution partners. For questions contact fieldcorrectiveaction@alignedmedicalsolutions.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2802-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology