Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Recall

Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Stryker Hoffmann II Connecting Rod Micro 3 x 40mm.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2828-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 1, 2026
FDA report date
Aug 5, 2026
Quantity in commerce
6,244 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.

Affected model & UDI checker

Search the exact identifiers published for Z-2828-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
50796040.Stryker Hoffmann II Connecting Rod Micro 3 x 40mm.Check source notice04546540389909
50796040Stryker Hoffmann II Connecting Rod Micro 3 x 40mm.Check source noticeNot specified
FDA code information

Catalog Number: 50796040. GTIN: 04546540389909. All lot numbers.

View the original FDA recall notice

Why was this product recalled?

Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did STRYKER GmbH take?

Firm began notifying consignees on July 1, 2026. Customers were instructed to immediately check their internal inventory to locate affected product. Customers were informed that until further notice, the subject devices are limited to single use only. Following first use, discard the devices in accordance with the facility's established procedures. Users should conduct visual inspection of product prior to each use. If cracking is found to be present, the customer should quarantine non-confirming parts and report the issue to Stryker's Trauma & Extremities division. Rods with cracks should be returned as part of the complaint process.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.