Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2843-2026

Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Recall

Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2843-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 25, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
1102
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Lot Code: Q6 Edge 3 Stretto 0060650920018*9 K143383 D246699 41875 ITI

View the original FDA recall notice

Why was this product recalled?

If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Pride Mobility Products Corp. take?

Firm notified dealers and consumers via mailed letters on June 25, 2026. Dealers were instructed to locate affected units and inform consumers of the required software update. Dealers should schedule service appointments with affected consumers. Systems should be updated to Q-Logic software version 2.0.13 or later. The update must be completed following a system power cycle. Consumers were instructed to disable the "Remind Me Later" feature immediately. Consumers should contact their dealer to schedule an appointment for software update.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology