Nerivio Infinity. Recall
Nerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.28.0.11
FDA recall Class II
Ongoing
FDA lists this Class II recall for Nerivio Infinity.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2855-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 23, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 11,886 units
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
Nerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.28.0.11
Affected model & UDI checker
Search the exact identifiers published for Z-2855-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| FGD000100 | Nerivio Infinity. | Check source notice | Not specified |
| Not specified | Nerivio Infinity. | Check source notice | 7290017578118 |
FDA code information
Model/Catalog Number: FGD000100. UDI: 01072900175781181125120221NI2544-0300673 Software Version: FW Version 4.1.2.0, App version 7.28.0.11 GTIN: 7290017578118. Serial Number: NI2544-0300673.
Why was this product recalled?
Potential for the batter to have a thermal-runway event during charging, resulting in smoke and odor.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did THERANICA BIOELECTRONICS LTD take?
Firm issued a stop ship of product from distributors starting March 22, 2026 after learning of the reported event. These devices were released from hold on 4/30/2026, after the device firmware update was released on 4/19/26. Starting May 13, 2026, customers received in-app messaging which ensure that all affected users receive instructions regarding the required corrective action. Users were prompted to perform a software update affecting both the mobile application used to control the device and the device firmware. The firmware update permanently disables the device upon detection of an overheating event. The software update provides an enhanced warning mechanism if it detects a device malfunction.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2855-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology