Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2858-2026

Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Recall

Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2858-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 21, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
3740 kits (180 cases)
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791

Affected model & UDI checker

Search the exact identifiers published for Z-2858-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
78791Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear SequenceCheck source noticeNot specified
Not specifiedBrand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear SequenceCheck source notice20809160242558
FDA code information

Lot Code: Lot No.: 327075 328690 319980 320965 318461 322375 325653 330775 331356 336680 UDI: 20809160242558

View the original FDA recall notice

Why was this product recalled?

Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medical Action Industries, Inc. 306 take?

On May 21, 2026 Medical Action Industries issued a "Urgent: Medical Device" Recall notification via Email. Medical Action Industries asked consignees to take the following actions: 1. Identify, Segregate, and Quarantine all impacted kits listed in Table 1. 2. Add warning labels to affected AVID kits to ensure all users are aware of discarding the impacted component prior to use. 3. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 4. Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6. Complete the attached Recall Response Form and return by June 5, 2026.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology