Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2875-2026

Merge OrthoPACS, medical imaging diagnostic workstation Recall

Merge OrthoPACS, medical imaging diagnostic workstation

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merge OrthoPACS, medical imaging diagnostic workstation. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2875-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 30, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
5 units
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Merge OrthoPACS, medical imaging diagnostic workstation

Affected model & UDI checker

Search the exact identifiers published for Z-2875-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMerge OrthoPACS, medical imaging diagnostic workstationCheck source notice00842000101154
FDA code information

UDI/DI (01)00842000101154(10)9.4.0.P02(11)260302, with Affected Versions: 9.4, 9.4 P01, and 9.4 P02.

View the original FDA recall notice

Why was this product recalled?

Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.

What actions did Merge Healthcare, Inc. take?

Merge Healthcare issued an Urgent Field Safety Notice (Revision 2) to its consignees on 6/30/2026 via email. The notice explained the issue and when it may occur, risk to patient, and requested an immediate workaround, a configuration change, be performed to disable automatic refresh of the Thumbnail Bar until a software update is available. The notice also directions the consignee to forward the notice to any individuals within their organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The notice was followed by telephone calls beginning on 7/6/2026 to ensure the notice was received. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology