GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system Recall
GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system
FDA recall Class II
Ongoing
FDA lists this Class II recall for GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2878-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 10, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 11 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system
Affected model & UDI checker
Search the exact identifiers published for Z-2878-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 5700592 | GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system | Check source notice | Not specified |
| Not specified | GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system | Check source notice | 00198953056752 |
FDA code information
GTIN 00198953056752, Lot/Serial Numbers: SWT25510001TA, 3M1D391215, 3M1D411081, 3M1D411083, 3M1D3B100D, SWT26110010TA, SWT26070001TA, SWT26120008TA, 3M1D411080, 3M1D411082, 3M1D3B100C.
Why was this product recalled?
A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information (PHI).
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Ge Medical Systems Information Technologies, Inc. take?
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 7/10/2026 via traceable means. The notice explained the safety issue, and requested the following: "Actions to be taken by Customer/User You can continue to use Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.2.0, pending GE HealthCare correction. If you need help to identify if your Portrait Mobile Monitoring Solution has Portrait Core Services SW 1.2.0, contact GE HealthCare Service. Ensure that your facility follows all hospital networking and cybersecurity policies and procedures along with the security considerations outlined in the GE HealthCare Portrait Mobile Monitoring Solution Privacy and Security Manual p/n 5970026. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or a local Service Representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2878-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology