Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2885-2026

Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Recall

Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2885-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 20, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
2
Distribution
See key details
FDA cause
Process design

What is being recalled?

Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-2885-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
9323-5333Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and rightCheck source noticeNot specified
9323-5334Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and rightCheck source noticeNot specified
Not specifiedBrand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and rightCheck source notice04719886908084
Not specifiedBrand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and rightCheck source notice04719886908091
FDA code information

Product Number; UDI-DI Code; Lot Code: Product number 9323-5333; UDI-DI 04719886908084; Lot number 25FB50 Product number 9323-5334; UDI-DI 04719886908091; Lot number 25FB51

View the original FDA recall notice

Why was this product recalled?

Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

What actions did UNITED ORTHOPEDIC CORP. take?

On 03/11/2026, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that due the fact that the PSA Box Trial cannot be properly assembled with the Hinge Trial in the affected devices because of incomplete processing of the screw components, which may result in interference during assembly with the corresponding device. On 05/20/2026, the firm emailed an updated "URGENT: MEDCIAL DEVICE RECALL" Letter Customers are instructed: 1. Examine your inventory immediately to locate the affected product whose quantities and lot numbers are listed above. Please note the product may have already been returned. 2. Cease all use of these devices immediately and quarantine all affected devices in your territory for immediate return. 3. Customer Service will be contacting you regarding replacements. If you still have the product, you are requested to immediately return the affected product using the prepaid airbills on RMA XXXX . a. Write the RMA number on the outside of the box clearly and legibly. b. Ship recalled product via Fed Ex standard overnight to: UOC USA INC. Attention: RMA# XXXX 16601 Hale Ave. Irvine, CA 92606 (Attached Shipping Label) 4. Additionally, if your firm has further distributed the recall product, please provide that customer with a copy of this recall notification and assist them with the return of the product. 5. Complete and return the following page according to the instructions. For Question - contact UOC Sr. Director of Sales Operations at 949.328.3366.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology