FDA RECALL RECORD · Z-2897-2026

Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mm Recall

Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mm Model/Catalog Number: 9323-5310 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mm. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2897-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 19, 2026
FDA report date
Aug 12, 2026
Quantity in commerce
3
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mm Model/Catalog Number: 9323-5310 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A

Affected model & UDI checker

Search the exact identifiers published for Z-2897-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
9323-5310Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mmCheck source noticeNot specified
Not specifiedBrand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mmCheck source notice04719886908060
FDA code information

Lot Code: Product Number 9323-5310; UDI-DI 04719886908060; Lot number 25FC58

View the original FDA recall notice

Why was this product recalled?

Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did UNITED ORTHOPEDIC CORP. take?

On 03/09/2026, the firm distributed an "URGENT: MEDICAL DEVICE RECALL" Letter informing customer that due the fact that the Femoral offset bushing in the affected devices is engraved with the incorrect numbering, and the rotational direction does not align with the intended orientation. Customers are instructed to: 1. Examine inventory immediately to locate the affected product whose quantities and lot numbers are listed in the letter. Note the product may have already been returned. 2. Cease all use of these devices immediately and quarantine all affected devices in their territory for immediate return. 3. Customer Service will be contacting customers regarding replacements. If customers still have the product, they are requested to immediately return the affected product using the prepaid airbills on RMA XXXX . a. Write the RMA number on the outside of the box clearly and legibly. b. Ship recalled product via Fed Ex standard overnight to: UOC USA INC. Attention: RMA# XXXX 16601 Hale Ave. Irvine, CA 92606 (Attached Shipping Label) 4. Additionally, if customers has further distributed the recall product, please provide that customer with a copy of this recall notification and assist them with the return of the product. 5. Complete and return the following page according to the instructions. For questions - contact UOC Sr. Director of Sales Operations at 949.328.3366.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.