Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF Recall
Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF
FDA recall Class II
Ongoing
FDA lists this Class II recall for Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2910-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 17, 2026
- FDA report date
- Aug 12, 2026
- Quantity in commerce
- 40 units
- Distribution
- See key details
- FDA cause
- Labeling mix-ups
What is being recalled?
Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF
Affected model & UDI checker
Search the exact identifiers published for Z-2910-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 1610-LF | Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF | Check source notice | Not specified |
| Not specified | Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF | Check source notice | 00813307014232 |
FDA code information
Lot #0000369205/UDI: 00813307014232
Why was this product recalled?
Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.
What actions did Freudenberg Medical, LLC take?
On June 26, 2026, Inhealth Technologies issued a Urgent Medical Device Recall Notification via FedEx to affected consignees. Inhealth ask consignees to take the following actions: 1. Do not use, ship, or distribute affected products 2. Identify and isolate all affected inventory 3. Destroy all affected devices 4. Complete the attached response form 5. Return confirmation no later than: July 8, 2026 6. Email confirmation of product destruction to: regulatoryaffairs@freudenbergmedical.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2910-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.