OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1 Recall
OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1
FDA recall Class II
Ongoing
FDA lists this Class II recall for OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2933-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 13, 2026
- FDA report date
- Aug 19, 2026
- Quantity in commerce
- 8,224 boxes (148,032 units)
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1
Affected model & UDI checker
Search the exact identifiers published for Z-2933-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| C020130-1 | OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1 | Check source notice | Not specified |
| Not specified | OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1 | Check source notice | 20888937027451 |
FDA code information
UDI: (01)20888937027451/ Lots: 24J1202Z, 25D2302Z, 25A0202Z, 25E1502Z, 25B2802Z, 25L0802Z
Why was this product recalled?
There is a potential for vaginal speculums to contain a burr on the proximal plastic blade, which may introduce a sharp edge in the contact regions for both the patient and the healthcare provider.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did OBP Medical Corporation take?
On July 14, 2026, OBP Medical a subsidiary of Cooper Surgical issued a Urgent Medical Device Recall Notification to affected consignees via UPS. OBP Medical asked consignees to take the following actions: 1. Physicians/Facilities: Please remove all product of the subject lots from your inventory and note the number of units to be returned on the attached Recall Response Form. 2. Distributors: Please remove all product of the subject lots from your inventory and note the number of units to be returned on the attached Distributor Only Acknowledgement Form. Per the instructions, please communicate this recall notification to all customers that you distributed these devices to. 3. If you are aware of malfunctions or adverse events involving the subject devices which you have not previously reported please report via: Cooper Surgical Customer Complaint Form or FDA's MedWatch Voluntary Form
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2933-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology