Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2938-2026

ERBECRYO 2 Cryosurgical Unit Recall

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

FDA recall Class II

Ongoing

FDA lists this Class II recall for ERBECRYO 2 Cryosurgical Unit. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2938-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 26, 2026
FDA report date
Aug 19, 2026
Quantity in commerce
1126 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Affected model & UDI checker

Search the exact identifiers published for Z-2938-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
10402-000ERBECRYO 2 Cryosurgical UnitCheck source noticeNot specified
Not specifiedERBECRYO 2 Cryosurgical UnitCheck source notice04050147013797
FDA code information

UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3

View the original FDA recall notice

Why was this product recalled?

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

What actions did Erbe USA Inc take?

On June 26, 2026, Erbe USA issued a Urgent: Medical Device Correction update notification to affected consignees via multiple methods ( Email/Website/Hand delivery). ERBE ask consignees to take the following actions: 1. We request the ERBECRYO 2 Unit be made available to the Erbe Representative to install the (V)1.0.4 Software. The Erbe Representative will perform all tasks required to update the software. 2. Confirm acknowledgement with Erbe Representative on-site. 3. Share this notice with all users within your facility. 4.While awaiting software update, there is no change of the Cryosurgical Unit s intended use or performance. Please continue to follow all warnings, instructions, guidelines, etc. involved with the Unit and Cryoprobe.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology