Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx. Recall

Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2951-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 10, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
32 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Software Version: Protocol Configuration Tool (PCT) v2.0.0; Equipment ID: 100071, 100213, 100355, 102281, 103047, 104154, 104201, 104202, 104204, 104205, 104224, 104503, 104504, 104562, 104885, 105961, 106497, 00016, 00487, 00488, 00489, 00490, 00817, 00491, 00492, 00854, 01173, 01174, 01785, 01337, 01596, 01993.

View the original FDA recall notice

Why was this product recalled?

Software issue that can cause an increase in spontaneous software closures.

What actions did Foundation Medicine, Inc. take?

An URGENT MEDICAL DEVICE RECALL NOTIFICATION dated 6/10/26 was emailed to consignees. Consignees were instructed to review the impacted equipment list and verify that the PCT V2.0.1 software version has been installed. If software hasn't been updated FMI will follow up to assist with installation. The recall notice is to be forwarded to anyone in the facility or outside the organization that needs to be informed. The notification is to be retained with laboratory records. Questions can be directed to Foundation Medicine at 888-988-3639 or QACTAdverseEventReporting@foundationmedicine.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.