Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2963-2026

GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge. Recall

GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.

FDA recall Class II

Ongoing

FDA lists this Class II recall for GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2963-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 30, 2026
FDA report date
Aug 26, 2026
Quantity in commerce
9 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.

Affected model & UDI checker

Search the exact identifiers published for Z-2963-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
6693662GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.Check source noticeNot specified
FDA code information

DUI/DI 07325710000304, Serial/Lot Numbers: 21049, 21088, 21089, 21093, 21100, 21103 21108, 21114, 21115.

View the original FDA recall notice

Why was this product recalled?

The module did not undergo electrical safety tests and functional tests during production.

What actions did Maquet Critical Care AB take?

Getinge issued an URGENT MEDICAL DEVICE CORRECTION notice was issued to its consignees on 7/9/2026 via FedEx. The notice explained the issue, potential risk to health, and requested the following: "Prior to the replacement, Getinge requests that you take the following actions: 1. Ensure all affected product in your inventory has been identified. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 3. Contact Technical Service if you have questions at 888-943-8872 (888 9GETUSA), Option 4, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). 4. Complete and return the enclosed response form to Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. For questions regarding this communication, please contact contact Technical Service if you have any questions at 888-943-8872 (888 9GETUSA), Option 4, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology