Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly Recall
Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly
FDA recall Class II
Ongoing
FDA lists this Class II recall for Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2965-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 13, 2026
- FDA report date
- Aug 26, 2026
- Quantity in commerce
- 24 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly
Affected model & UDI checker
Search the exact identifiers published for Z-2965-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly | Check source notice | 00630414976891 |
FDA code information
Lot Code: D387/UDI: 00630414976891
Why was this product recalled?
Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive.
What actions did Siemens Healthcare Diagnostics Products, Ltd. take?
On July 13, 2026, Siemens Healthineers issued a Urgent Medical Device Correction to affected consignees via FedEx. Siemens asked consignees to take the following actions: 1. Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2.Discontinue use and discard the affected kit lot. 3. Review your current inventory to determine your laboratory s replacement needs and provide this information to Siemens Healthineers for reporting to the authorities. 4. Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. 5.Retain this letter with your laboratory records and forward this letter to those who may have received the affected product(s).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2965-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026