Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001 Recall
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
FDA recall Class II
Ongoing
FDA lists this Class II recall for Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2967-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 20, 2026
- FDA report date
- Aug 26, 2026
- Quantity in commerce
- 1445 systems
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Affected model & UDI checker
Search the exact identifiers published for Z-2967-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| PRGNX-8008-001 | Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001 | Check source notice | Not specified |
| Not specified | Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001 | Check source notice | 00850005470342 |
FDA code information
All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2
Why was this product recalled?
Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
What actions did Paragonix Technologies Inc. take?
On July 20, 2026, Paragonix Technologies Inc./Getinge issued an Urgent Medical Device Correction to affected consignees via FedEX mail. Paragonix requested consignees to take the following actions: 1. An update to the Paragonix App has been released through the application store. 2. We recommend updating all mobile devices to latest version prior to KidneyVault usage. 3. Please continue to inspect all KidneyVault devices per the IFU before use to ensure proper datalogger function. 4. Continue to follow the IFU for using a multi-faceted clinical approach to assessing the organ, including any/all information available on the datalogger screen. 5. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 6. Complete and return the enclosed response form by e-mailing a copy to recallresponses.qrc@getinge.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2967-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology