Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2988-2026

Philips Incisive CT with software V4.5&V5.0, Product Code 728143 and Product Code 728144. Recall

Philips Incisive CT with software V4.5&V5.0, Product Code 728143 and Product Code 728144.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips Incisive CT with software V4.5&V5.0, Product Code 728143 and Product Code 728144.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2988-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 26, 2026
FDA report date
Sep 2, 2026
Quantity in commerce
N/A
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

Philips Incisive CT with software V4.5&V5.0, Product Code 728143 and Product Code 728144.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Device Identifier: (01)00884838085015, (01)00884838105508.

View the original FDA recall notice

Why was this product recalled?

Potential of failing to trigger clinical scan during Bolus Tracking.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips Healthcare (Suzhou) Co., Ltd. take?

Philips notification dated June 26, 2026 will inform customers of the issue via an Electronic Product Radiation Correction notification. Philips will schedule an appointment with customers to update the software free of charge. Customers who need additional information or support concerning this issue should be directed to their local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center). Philips reference: 2026-PD-CTAMI-108.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology