Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation Recall
Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation
FDA recall Class II
Ongoing
FDA lists this Class II recall for Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3004-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 22, 2026
- FDA report date
- Sep 2, 2026
- Quantity in commerce
- 20 units
- Distribution
- See key details
- FDA cause
- Software Design Change
What is being recalled?
Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation
Affected model & UDI checker
Search the exact identifiers published for Z-3004-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation | Check source notice | 00842000101239 |
FDA code information
Software version V9.5 - UDI (01)00842000101239(10)9.5.0(11)260409
Why was this product recalled?
Software issue may cause measurements to be displayed inaccurately.
What actions did Merge Healthcare, Inc. take?
Merge Healthcare issued an Urgent Field Safety Notice to its consignees on 7/22/2026 via email with a follow-up telephone call. The notice explained the issue and how to identify it, risk to patient, and requested the following: "Immediate actions to reduce risk to patients Merge Healthcare recommends that customers contact Merge Healthcare Support to identify any affected ultrasound studies and the necessary next steps to be taken. Until the software update is installed, ultrasound measurements performed using the Merge PACS & OrthoPACS Viewer should be confirmed against the measurements acquired at the ultrasound modality." The required actions are as follows: "Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Assist you in identifying and reviewing any ultrasound studies with measurements that may have been affected. iii) Schedule the installation of the software update that eliminates the root cause of this issue. Please forward this notice to any individuals within the organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3004-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology