FDA RECALL RECORD · Z-3013-2026

Midmark Digital ECG HD+12, HD+15 Recall

Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Midmark Digital ECG HD+12, HD+15. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3013-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 20, 2026
FDA report date
Sep 2, 2026
Quantity in commerce
3624 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

Affected model & UDI checker

Search the exact identifiers published for Z-3013-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
4-000-0080Midmark Digital ECG HD+12, HD+15Check source noticeNot specified
Not specifiedMidmark Digital ECG HD+12, HD+15Check source notice00841709121319
Not specifiedMidmark Digital ECG HD+12, HD+15Check source notice08056732652491
FDA code information

Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark Digital ECG. The Midmark Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 7817519 7817645, L1130587 V2967564 UDI-DI: 00841709121319, 08056732652491

View the original FDA recall notice

Why was this product recalled?

A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Midmark Corporation take?

n URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/17/26 was sent to customers. Safety Instructions Until the corrected IQecg Plugin software has been installed, for electrocardiogram tracings that reflect a fast heart rate (tachycardia), Midmark recommends performing a manual clinical assessment of the patient's heart rate prior to initiating therapeutic intervention. This bedside verification should be used to confirm the accuracy of the displayed ECG heart rate before making clinical management decisions based on the device's displayed value. If this issue is encountered, follow the temporary workaround instructions provided in Appendix A. Midmark will contact your facility with additional information regarding verification of your software installation and implementation of this Field Corrective Action. Product Correction Midmark has developed a corrected version of the IQecg Plugin software that addresses the issue described in this notification and is implementing a Field Corrective Action for affected software versions. No immediate action is required to contact Midmark at this time. Midmark will contact your facility to verify the IQecg Plugin software version(s) installed on your workstations and coordinate the appropriate correction process. Depending on your facility's configuration and support requirements, the correction may be completed using Midmark-provided installation instructions or with assistance coordinated by Midmark Technical Support or an authorized service provider. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email techsupport@midmark.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.