Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3030-2026

FUJIFILM Synapse PACS software v7.4.310; radiological image processing system Recall

FUJIFILM Synapse PACS software v7.4.310; radiological image processing system

FDA recall Class II

Ongoing

FDA lists this Class II recall for FUJIFILM Synapse PACS software v7.4.310; radiological image processing system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3030-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 22, 2026
FDA report date
Sep 2, 2026
Quantity in commerce
6 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

FUJIFILM Synapse PACS software v7.4.310; radiological image processing system

Affected model & UDI checker

Search the exact identifiers published for Z-3030-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedFUJIFILM Synapse PACS software v7.4.310; radiological image processing systemCheck source notice00854904006008
FDA code information

UDI (01)00854904006008(10)0704310, Serial Numbers: CV-00272120, CV-0080080176, CV-01318490, CV-01-80102732, CV-80057523, CV-00494020-38

View the original FDA recall notice

Why was this product recalled?

Software issue may cause the ruler measurement to be inaccurate.

What actions did FUJIFILM Healthcare Americas Corporation take?

FUJIFILM issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 7/21/2026 via email. The notice explained the issue and how to identify it, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1. Please post this notification on or near the Synapse PACs workstation. 2. Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 3. Please contact Technical Assistance Center (TAC) at 1-888-FUJIMED (1-888-385-4633) to open a service ticket. 4. Please report any known adverse events to FDA via MedWatch. Regarding this issue, Synapse PACS can safely be used by the end-user as long as: 5. The end-user is aware of the issue and, 6. Appropriate actions are taken when rendering a diagnosis for patients. The following workarounds can be used: 7. Perform M-mode ruler measurements on the ultrasound system and then transfer the study to Synapse PACS." For questions regarding this notice, please email hcusquality@fujifilm.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.