healthmark ProChek-W Protein Test for Robotic Arms; ProChek-Semi Quantitative Protein Test, Model Number PT-SQ-001 Recall
healthmark ProChek-W Protein Test for Robotic Arms; ProChek-Semi Quantitative Protein Test, Model Number PT-SQ-001
FDA recall Class II
Ongoing
FDA lists this Class II recall for healthmark ProChek-W Protein Test for Robotic Arms; ProChek-Semi Quantitative Protein Test, Model Number PT-SQ-001. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3032-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 29, 2026
- FDA report date
- Sep 2, 2026
- Quantity in commerce
- 53 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
healthmark ProChek-W Protein Test for Robotic Arms; ProChek-Semi Quantitative Protein Test, Model Number PT-SQ-001
Affected model & UDI checker
Search the exact identifiers published for Z-3032-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| PT-SQ-001 | healthmark ProChek-W Protein Test for Robotic Arms; ProChek-Semi Quantitative Protein Test, Model Number PT-SQ-001 | Check source notice | Not specified |
FDA code information
Serial/Lot Numbers: 200035, 200036, 200037, 200038, 200039, 200040, 30028.
Why was this product recalled?
Test manufactured with an expired ingredient which may present as either a false negative or positive.
What actions did Healthmark Industries Co., Inc. take?
GETINGE issued an URGENT MEDICAL DEVICE RECALL/REMOVAL to its consignees on 7/29/2026 via email and telephone. The notice explained the issue, risks to health, and requested the following: "Customer Actions Our records indicate that your facility has received one or more of the affected devices that are affected by this notice. As a result, Healthmark requests that your facility takes the following actions: 1. Ensure all affected product in your facility s inventory is no longer in use and has been properly quarantined. 2. Please ensure this message is forwarded to any individuals who need notification within your organization or any organization where the affected devices have been transferred. 3. Contact Customer Service via email at HMCS@hmark.com or 800-521-6224 to return your unused affected product or if you have any questions. 4. Please fill out the Customer Response Form using the link or QR code provided below. Getinge has partnered with IQVIA MedTech to coordinate the recall response process."
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3032-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology