FDA RECALL RECORD · Z-3033-2026

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000 Recall

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

FDA recall Class II

Ongoing

FDA lists this Class II recall for ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3033-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 6, 2026
FDA report date
Sep 2, 2026
Quantity in commerce
6 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

Affected model & UDI checker

Search the exact identifiers published for Z-3033-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000Check source notice00857534006349
FDA code information

113821-600 - Serial Numbers: 10003754, 10006997; 114148-000 - UDI-DI: 00857534006349, Serial Numbers: 10005189, 10005190, 10004463, 10004997

View the original FDA recall notice

Why was this product recalled?

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

What actions did IMRIS Imaging Inc take?

An URGENT: MEDICAL DEVICE RECALL notification letter dated 8/6/26 was sent to customers. IMRIS recommends the following preventative actions until the corrective actions have been implemented on your table: " Per the OR Table Operator Manuals, the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, tabletops, headrest, pendant connection and rotation. " During usage, verify the table position is locked by manually pushing the table to ensure it does not exhibit rotational movement. " If the floor latching mechanism is observed to not function properly, please contact IMRIS Customer Service immediately. Please confirm receipt of this notice with IMRIS using the Customer Acknowledgement Form below, and please forward this communication to any others inside or outside your organization who may work with the affected ORT200 and ORT300 tables. To minimize the potential risk of injury, IMRIS is taking the following corrective actions: " The existing ORT200 and ORT300 linear actuator and rotational lock mechanism components will be replaced with an updated design. Our customer service organization will contact you to make arrangements for IMRIS field service personnel to perform the correction at your facility. IMRIS has identified that your facility has one or more ORT200 or ORT300 Operating Tables which are subject to this recall notice. Please confirm details on the last page of this letter and return your response to IMRIS. IMRIS Customer Service will contact you to facilitate replacement of the linear actuator and rotational lock mechanism components at your facility. Please contact our customer support team through our service line at 1-866-475-0525 or e-mail: customersupport@imris.com if you have any questions or concerns. Please keep this information until the corrective measures have been finalized.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology