HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU) Recall
HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU)
FDA recall Class II
Ongoing
FDA lists this Class II recall for HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3064-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 31, 2026
- FDA report date
- Sep 9, 2026
- Quantity in commerce
- 111,478
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU)
Affected model & UDI checker
Search the exact identifiers published for Z-3064-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 10300012 | HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU) | Check source notice | Not specified |
| 10300016 | HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU) | Check source notice | Not specified |
| 10301012 | HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU) | Check source notice | Not specified |
| 10301016 | HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU) | Check source notice | Not specified |
FDA code information
REF/UDI-DI/Lot: 10300012/5050474755093/UA32232, UA32250, UA32268, UA32290, UA32292, UA32303, UA32371, UP31203, UP31205; 10300016/5050474755116/UA32447, UA32448, UA32449, UB33008, UB33026, UB33082, UB33083, UB33175; 10301012/5050474755109/UB33014, UB33041, UB33057, UB33060, UB33100, UB33190; 10301016/5050474755123/UB33084, UB33194
Why was this product recalled?
Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
What actions did Johnson & Johnson Surgical Vision, Inc. take?
On 7/31/2026, correction notices were mailed to customers and distributors informing them of the following: Device Insertion Technique Clarification: If resistance is encountered during insertion, withdraw the cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding. Do not force the cannula through the incision, as this may increase the risk of unintended tissue injury, including, but not limited to corneal abrasion, iris trauma or Descemet s Membrane Detachment, which may require medical or surgical intervention to preclude serious injury. This investigation is currently ongoing, and we will notify users if any additional information comes to light. 1) Forward this notice to all surgeons using this device in your facility. 2) If any of the subject product has been transported to another facility/surgeon, contact that facility/surgeon and provide/forward this notice to them. 3) Keep a copy of this notice with the subject product. 4) Complete, sign, and return the Business Reply Form via email to RegCompliOne@its.jnj.com If you have questions or concerns with regard to this notification, please contact the firm's Global Post Market Safety Team at 1-877-266-4543, Option 2. Additional notices with additional affected devices to be sent in the future.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3064-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology