Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3082-2026

LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7) Recall

LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7)

FDA recall Class II

Ongoing

FDA lists this Class II recall for LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3082-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 6, 2026
FDA report date
Sep 9, 2026
Quantity in commerce
N/A
Distribution
See key details
FDA cause
Software design

What is being recalled?

LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7)

Affected model & UDI checker

Search the exact identifiers published for Z-3082-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
7246LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7)Check source noticeNot specified
Not specifiedLUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7)Check source notice00802526613883
FDA code information

UDI-DI: 00802526613883

View the original FDA recall notice

Why was this product recalled?

There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did BOSTON SCIENTIFIC CORPORATION take?

An Urgent Medical Device Correction dated August 2026 was sent to customers. Boston Scientific is notifying you of the potential for a delay in AF Burden Alert notification affecting the LATITUDE" Clarity server software which supports the LUX-Dx" family of ICMs. This delayed AF Burden Alert notification behavior by LATITUDE Clarity may occur if the Alert for AF Burden e Threshold is configured to ">0 hours per day" setting (a non-default configuration). The ICM is performing as intended with continued AF detection, data storage, and transmission of device data. Boston Scientific has received one complaint related to this delayed notification behavior by the LATITUDE Clarity System. No patient injuries or deaths have been reported. A server software update is planned to correct this behavior. Configuration Options for Atrial Alert Notifications: Atrial arrhythmia alert notification preferences may vary, and AF configurations may be set for individual patients and/or group settings. Until updated server software is available, consider the following options: Change Threshold for AF Burden Alert Setting Enable AF Events Alert with Minimum Duration If Unchanged, Potential Delay in AF Burden Alert Notification " Boston Scientific does not recommend replacement of the implanted ICM. " Please share this notification with all HCPs within your organization who manage patients using the LUXDx ICM and LATITUDE Clarity System. Mitigation. The LATITUDE Clarity system is designed to detect atrial fibrillation through two independent mechanisms: AF Burden and AF Event Alerts. Until a software update to the LATITUDE Clarity server software is available, configuring AF Events or changing AF Burden Alerts settings can mitigate delays in full device interrogation and enable the LATITUDE Clarity system to display alert notification as intended. Additional Information. The Regulatory Authority of your country has been informed of this communication. Adverse events should be reported

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology