5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3 Recall
5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3
FDA recall Class II
Ongoing
FDA lists this Class II recall for 5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3085-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 10, 2026
- FDA report date
- Sep 9, 2026
- Quantity in commerce
- 62,544 units
- Distribution
- See key details
- FDA cause
- Component change control
What is being recalled?
5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3
Affected model & UDI checker
Search the exact identifiers published for Z-3085-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | 5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3 | Check source notice | 00840861102518 |
FDA code information
Lots 25JR01800 25KR01224 25LR01001 25NR01018 25NR01272 25PR01077 25PR01219 25PR01220 UDI-DI 00840861102518
Why was this product recalled?
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
What actions did Fresenius Medical Care Renal Therapies Group, LLC take?
On 8/10/2026, the firm began notifying consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to 1. Identify any affected product on hand. 2. IMMEDIATELY segregate and discontinue use of affected lot. 3. Contact the call center at 1-888-943-4901 and reference Event 8514, between 8:00 AM 4:30 PM Eastern daily, for return of affected lots and replacement of your stock. 4, Complete and return the included reply form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3085-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.