Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-3112-2026

Getinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136); Recall

Getinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);

FDA recall Class II

Ongoing

FDA lists this Class II recall for Getinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3112-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 7, 2026
FDA report date
Sep 16, 2026
Quantity in commerce
2533
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Getinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);

Affected model & UDI checker

Search the exact identifiers published for Z-3112-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGetinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);Check source notice00650862132243
Not specifiedGetinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);Check source notice00650862132281
Not specifiedGetinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);Check source notice00650862132328
Not specifiedGetinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2. 28FR(8128), 3. 32FR(8132), 4. 36FR(8136);Check source notice00650862132366
FDA code information

1. Model Number(Size): 8124(24FR); UDI-DI: 00650862132243; Lot Number: ME244852; 2. Model Number(Size): 8128(28FR); UDI-DI: 00650862132281; Lot Number: ME244854, ME244855; 3. Model Number(Size): 8132(32FR); UDI-DI: 00650862132328; Lot Numbers: ME244565, ME244858, ME244859; 4. Model Number(Size): 8136(36FR); UDI-DI: 00650862132366; Lot Number: ME244861;

View the original FDA recall notice

Why was this product recalled?

An Out of Specification result (OOS) related to bacterial endotoxin testing in a single lot of catheters.

What actions did Atrium Medical Corporation take?

On August 7, 2026 URGENT MEDICAL DEVICE - RECALL letters were sent to customers. Customer Actions: 1. Please examine your inventory immediately to determine if you have any of the Atrium PVC Thoracic Catheters with the specific lot numbers. 2. Should you have any affected product, please remove from area(s) of use and quarantine. 3. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 4. Please contact Contact Customer Service via email at customersupport.us@getinge.com or 888-943-8872, option 2 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone) to request a return authorization (RMA) and shipping instructions to return any affected product. 5. Whether or not you have affected product(s) within expiration listed in this notice, please complete and sign the attached MEDICAL DEVICE- REMOVAL RESPONSE FORM to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology