Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333 Recall

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3155-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 18, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
62
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Affected model & UDI checker

Search the exact identifiers published for Z-3155-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333Check source notice00884838101111
FDA code information

UDI: (01)00884838101111

View the original FDA recall notice

Why was this product recalled?

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America Llc take?

Philips sent an Electronic Product Radiation Correction notification dated August 18, 2026 to the users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will schedule an appointment with customers to install a software solution free of charge.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.