Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333 Recall
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3155-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 18, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 62
- Distribution
- See key details
- FDA cause
- Radiation Control for Health and Safety Act
What is being recalled?
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
Affected model & UDI checker
Search the exact identifiers published for Z-3155-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333 | Check source notice | 00884838101111 |
FDA code information
UDI: (01)00884838101111
Why was this product recalled?
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips North America Llc take?
Philips sent an Electronic Product Radiation Correction notification dated August 18, 2026 to the users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will schedule an appointment with customers to install a software solution free of charge.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3155-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology