Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Recall
Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3216-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 19, 2026
- FDA report date
- Sep 30, 2026
- Quantity in commerce
- 305 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355.
Affected model & UDI checker
Search the exact identifiers published for Z-3216-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| BD-420X-1355 | Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. | Check source notice | Not specified |
| Not specified | Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. | Check source notice | 00821925042261 |
FDA code information
Model Number: BD-420X-1355. UDI-DI: 00821925042261. Lot Number: 460898.
Why was this product recalled?
Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Olympus Corporation of the Americas take?
Olympus notified their direct customer on about 08/19/2026 via emailed "URGENT: MEDICAL DEVICE RECALL/PRODUCT REMOVAL" letter. The consignee was instructed to examine inventory for affected units, acknowledge receipt of the recall notification by completing and returning the Reply Form, arrange for the return of affected units for replacement, and if affected units have been further distributed, notify those customers.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3216-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology